The country of origin for this product is the United States. To see the full specifications for this product, please see our product manual found under the documentation tab if available
Thermocouple
Type | Range | Accuracy | Resolution |
J | -210° to 760°C (-346° to 1,400°F) | ±0.6°C (±1.08°F) ± thermocouple probe accuracy | 0.03°C (0.06°F) |
K | -260° to 1,370°C (-436° to 2,498°F) | ±0.7°C (±1.26°F) ± thermocouple probe accuracy | 0.04°C (0.07°F) |
T | -260° to 400°C (-436° to 752°F) | ±0.6°C (±1.08°F) ± thermocouple probe accuracy | 0.02°C (0.03°F) |
E | -260° to 950°C (-436° to 1,742°F) | ±0.6°C (±1.08°F) ± thermocouple probe accuracy | 0.03°C at (0.05°F) |
R | -50° to 1,550°C (-58° to 2,822°F) | ±2.2°C (±3.96°F) ± thermocouple probe accuracy | 0.08°C (0.15°F) |
S | -50° to 1,720°C (-58° to 3,128°F) | ±2.2°C (±3.96°F) ± thermocouple probe accuracy | 0.08°C (0.15°F) |
B | 550° to 1,820°C (1,022° to 3,308°F) | ±2.5°C (±4.5°F) ± thermocouple probe accuracy | 0.1°C (0.18F) |
N | -260° to 1,300°C (-436° to 2,372°F) | ±1.0°C (±1.8°F) ± thermocouple probe accuracy | 0.06°C (0.11°F) |
Internal 10K Thermistor (Temperature):
Range: Logging: -20° to 70°C (-4° to 158°F)
Acuracy: ±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A
Resolution: 0.024°C at 25°C (0.04°F at 77°F); see Plot A
Drift: <0.1°C (0.18°F) per year
Logger
Logger Operating Range
Logging: -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing); logger accuracy from 0° to 50°C (32° to 122°F)
Launch/readout: 0° to 50°C (32° to 122°F) per USB specification
Logging rate: 1 second to 18 hours, 12 minutes, 15 seconds
Logging modes: Normal, Burst or Statistics
Memory modes: Wrap when Full or Stop when Full
Start modes: Immediately, Push Button, Date & Time or Next Interval
Stop modes: When Memory Full, Push Button, or Date & Time
Restart mode: Push Button
Time accuracy: ±1 minute per month at 25°C (77°F), see Plot B
Battery life: 1 year, typical with logging rate of 1 minute and sampling interval of 15 seconds or greater
Battery type: Two AAA 1.5V alkaline batteries, user replaceable
Memory: 4 MB (1.6 million measurements, maximum)
Download type: USB 2.0 interface
Full memory download time: Approximately 1.5 minutes
LCD: LCD is visible from: 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
Size: 10.8 x 5.41 x 2.54 cm (4.25 x 2.13 x 1 in.)
Weight: 107.5 g (3.79 oz)
Environmental rating: IP50
- The CE Marking identifies this product as complying with all relevant directives in the European Union (EU)